Products

Makrolon® 2658

Makrolon® 2658

Manufacturer: Covestro
Produced At:
 Europe / China
Product Category: Polycarbonate
PC
Product No: Makrolon 2658 
Product Quantity: 25KG / BAG
Product Features: Biocompatible、Ethylene Oxide Sterilizable、Good Mold Release、Medium Viscosity、Steam Sterilizable

Makrolon® 2658

Polycarbonate
Product Description:
MVR (300 °C/1.2 kg) 12 cm³/10 min; medical devices; suitable for ETO and steam sterilization at 121 °C; biocompatible according to many ISO 10993-1 test requirements; medium viscosity; easy release; injection molding - melt temperature 280 - 320 °C; available in transparent and opaque colors

Features 
  • Biocompatible,,Ethylene Oxide Sterilizable,

  • Good Mold Release,Medium Viscosity,Steam Sterilizable

Uses 
  • Medical Devices,Medical/Healthcare Applications 

Agency Ratings 
  • ISO 10993-Part 1

RoHS Compliance 
  • RoHS Compliant

Appearance 
  • Clear/Transparent,Colors Available,Opaque

Processing Method 
  • Injection Molding

Multi-Point Data 
  • Creep Modulus vs. Time (ISO 11403-1)

  • Isochronous Stress vs. Strain (ISO 11403-1)

  • Isothermal Stress vs. Strain (ISO 11403-1)

  • Secant Modulus vs. Strain (ISO 11403-1)

  • Specific Volume vs Temperature (ISO 11403-2)

  • Viscosity vs. Shear Rate (ISO 11403-2)

Physical
Nominal Value
Unit
Test Method
Density (73°F)
1.20
g/cm³
ISO 1183
Apparent (Bulk) Density 3
0.66
g/cm³
ISO 60
Melt Mass-Flow Rate (MFR) (300°C/1.2 kg)
13
g/10 min
ISO 1133
Melt Volume-Flow Rate (MVR) (300°C/1.2 kg)
0.732
in³/10min
ISO 1133
Molding Shrinkage



Across Flow
0.60 to 0.80
%
ISO 2577
Flow
0.60 to 0.80
%
ISO 2577
Across Flow : 0.0787 in 4
0.75
%
ISO 294-4
Flow : 0.0787 in 4
0.70
%
ISO 294-4
Water Absorption


ISO 62
Saturation, 73°F
0.30
%

Equilibrium, 73°F, 50% RH
0.12
%

Hardness
Nominal Value
Unit
Test Method
Ball Indentation Hardness
16700
psi
ISO 2039-1
Mechanical
Nominal Value
Unit
Test Method
Tensile Modulus (73°F)
348000
psi
ISO 527-2/1
Tensile Stress


ISO 527-2/50
Yield, 73°F
9570
psi

Break, 73°F
10200
psi

Tensile Strain


ISO 527-2/50
Yield, 73°F
6.1
%

Break, 73°F
130
%

Nominal Tensile Strain at Break (73°F)
> 50
%
ISO 527-2/50
Tensile Creep Modulus


ISO 899-1
1 hr
319000
psi

1000 hr
276000
psi

Flexural Modulus 5 (73°F)
348000
psi
ISO 178
Flexural Stress 5


ISO 178
73°F
14100
psi

3.5% Strain, 73°F
10600
psi

Flexural Strain at Flexural Strength 6 (73°F)
7.1
%
ISO 178
Impact
Nominal Value
Unit
Test Method
Charpy Notched Impact Strength 7


ISO 7391
-22°F, Complete Break
7.6
ft·lb/in²

73°F, Partial Break
33
ft·lb/in²

Charpy Unnotched Impact Strength


ISO 179/1eU
-76°F
No Break


-22°F
No Break


73°F
No Break


Notched Izod Impact Strength 8


ISO 7391
-22°F, Complete Break
7.1
ft·lb/in²

73°F, Partial Break
33
ft·lb/in²

Multi-Axial Instrumented Impact Energy


ISO 6603-2
-22°F
47.9
ft·lb

73°F
44.3
ft·lb

Multi-Axial Instrumented Impact Peak Force


ISO 6603-2
-22°F
1420
lbf

73°F
1210
lbf

Thermal
Nominal Value
Unit
Test Method
Heat Deflection Temperature



66 psi, Unannealed
279
°F
ISO 75-2/B
264 psi, Unannealed
255
°F
ISO 75-2/A
Glass Transition Temperature 9
293
°F
ISO 11357-2
Vicat Softening Temperature



--
291
°F
ISO 306/B50
--
293
°F
ISO 306/B120
Ball Pressure Test (277°F)
Pass

IEC 60695-10-2
CLTE


ISO 11359-2
Flow : 73 to 131°F
3.6E-5
in/in/°F

Transverse : 73 to 131°F
3.6E-5
in/in/°F

Thermal Conductivity 10 (73°F)
1.4
Btu·in/hr/ft²/°F
ISO 8302


Products are widely used in plastic modification, automotive, construction, electrical and electronic, industrial, daily necessities and other fields


Consultation
展开